Nonin Onyx Vantage Finger Pulse Oximeter
Nonin Onyx Vantage Finger Pulse Oximeters are small, lightweight, portable devices indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly perfused.
Nonin Onyx Vantage Finger Pulse Oximeters Use
Nonin Onyx Vantage Finger Pulse Oximeters are intended for spot-checking of adult and pediatric patients on digits (fingers, thumb, toes) in a wide range of environments to include hospitals, clinics, long-term care facilities, skilled nursing facilities, emergency medical services, and home healthcare services.
Nonin Onyx Vantage Finger Pulse Oximeter Specifications:
Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Accuracy Summary by Decade – Finger and Thumb
Decade Oxygen Saturation (Arms)
70 – 80% ±2
80 – 90% ±2
90 – 100% ±2
70 – 100% ±2
This graph shows plots of the error (SpO2 – SaO2)
by SaO2 using the 9590 with a linear regression fit
and upper 95% and lower 95% limits of agreement.
Each sample data point is identified by the subject
from a clinical study in non-motion conditions.
Accuracy Summary by Decade – Toe
Decade Oxygen Saturation (Arms)
70 – 80% ±2
80 – 90% ±3
90 – 100% ±3
70 – 100% ±3
This graph shows plots of the error (SpO2 – SaO2) by SaO2 using the 9590 with a linear regression fit and upper 95% and lowers 95% limits of agreement. Each sample data point is identified by the subject from a clinical study using toes in non-motion conditions.
SpO2 Low Perfusion Accuracy (Arms*) 70 to 100% ±2 digits
Pulse Rate Declared Accuracy Range (Arms*): 20 to 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy Range (Arms*): 40 to 240 BPM ±3 digits
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW maximum average
Infrared: 910 nanometers @ 1.2 mW maximum average
Temperature:
Operating: 23 °F to 104 °F / -5 °C to 40 °C
Storage/Transportation: -40 °F to 158 °F / -40 °C to 70 °C
Time (from storage) for the monitor to be ready for its intended use: 3 minutes to warm from -40 °C to -5 °C
5 minutes to cool from 70 °C to 40 °C
Humidity:
Operating: 10% to 90% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Altitude:
Operating: Up to 13,123 feet / 4,000 meters
Hyperbaric Pressure: Up to 4 atmospheres
Battery Life:
Operating: Approximately 6,000 spot checks or 36 hours of continuous operation
using new alkaline batteries.
Storage: 12 months
Classifications per ANSI/AAMI ES60601-1 / CAN/CSA-C22.2 No. 60601-1:
Degree of Protection: Type BF-Applied Part
Enclosure Degree of Ingress Protection: IP32
Mode of Operation: Continuous
This product complies with ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing.
This device is not made with natural rubber latex.
*± 1 Arms represents approximately 68% of measurements at zero bias.